Mechanica Micro pipette, Single & Multi-fluvium, Novi ac Fixum Volumen
MicroPette plus
MicroPette plus multi alveo
Features
• dispensatio circumagatur pro illaboriosa pipetting commodum
• Singulare piston et conum conum facilem refectionem et sustentationem permittunt
• Compositum materia facta tip pyramidis munit altum signationem perficientur
• Compatible cum maxime universalis tip brands
• Online calibration is available
plene autoclavable
MicroPette Plus pipettae plene autoclavabiles sunt, facile purgatio et periculum contaminationis minuit.Vapor autoclaving perfici potest in 121°C, 1 talea pro 20 minutis.Post autoclavationem, fistula refrigerari debet et siccari per XII horas ante usum.
Commendatur ad inhibenda fistulae faciendae post singulas autoclavationes.
Gringo et obsignando pistonium pipettum post singulas 10th autoclavationes consequetur munus auctum.
MicroPette
Features
• Pipettes involucrum volumen ab 0.1μL ad 10mL
• Ergonomic consilio optimum operating experientia praebet
• Large propono fenestra permittit facile volumen identificatio
• Securus calibratiis et sustentationem
• Quisque MicroPette cum singulis libellum calibratiis secundum ISO8655
• Online calibration is available
MicroPette multi-fluvium
Features
• 8 et 12 fistulae canalis aptae sunt ad 96 laminas bene
• dispensatio circumagatur pro illaboriosa pipetting commodum
• Singulare piston et apice pyramidum ecclesiarum facile reparationem et sustentationem permittens
• Compone materia facta tip pyramidis se- cures altae signationis effectus
• Compatible cum maxime universalis tip brands
• Online calibration is available
Calibration
Omnes pipettae qualitates secundum ISO8655-2:2002 probatae sunt et singulis testimonialibus calibratiis praebentur.Quale imperium includit probatio gravimetrica uniuscuiusque fistulae cum aqua decocta in 22°C Nostra website permittit utentes utentes accessum programmatis calibrationis online et calibrationem accuratam et opportune consequantur.
Online calibratiis software Free ad pipette users
Specifications
Volume Range | Incrementum | Test Volume(μl) | accurate error | Subtilitas errorum | ||
% | μl | % | μl | |||
0.1-2.5μl | 0.05μl | 2.5 | 2.50% | 0.093 | 2.00% | 0.05 |
1.25 | 3.00% | 0.093 | 3.00% | 0.093 | ||
0.25 | 12.00% | 0.03 | 6.00% | 0.014 | ||
0.5-10μl | 0.1μl | 10 | 1.00% | 0.1 | 0.80% | 0.08 |
5 | 1.50% | 0.093 | 1.50% | 0.093 | ||
1 | 2.50% | 0.024 | 1.50% | 0.014 | ||
2-20μl | 0.5μl | 20 | 0.90% | 0.18 | 0.40% | 0.08 |
10 | 1.20% | 0.12 | 1.00% | 0.1 | ||
2 | 3.00% | 0.06 | 2.00% | 0.04 | ||
5-50μl | 0.5μl | 50 | 0.60% | 0.3 | 0.30% | 0.15 |
25 | 0.90% | 0.225 | 0.60% | 0.15 | ||
5 | 2.00% | 0.1 | 2.00% | 0.1 | ||
10-100μl | 1μl | 100 | 0.80% | 0.8 | 0.15% | 0.15 |
50 | 1.00% | 0.5 | 0.40% | 0.2 | ||
10 | 3.00% | 0.3 | 1.50% | 0.15 | ||
20-200μl | 1μl | 200 | 0.60% | 1.2 | 0.15% | 0.3 |
100 | 0.80% | 0.8 | 0.30% | 0.3 | ||
20 | 3.00% | 0.6 | 1.00% | 0.2 | ||
50-200μl | 1μl | 200 | 0.60% | 1.2 | 0.15% | 0.3 |
100 | 0.80% | 0.8 | 0.30% | 0.3 | ||
50 | 1.00% | 0.5 | 0.40% | 0.2 | ||
100-1000μl | 5μl | 1000 | 0.60% | 6 | 0.20% | 2 |
500 | 0.70% | 3.5 | 0.25% | 1.25 | ||
100 | 2.00% | 2 | 0.70% | 0.7 | ||
200-1000μl | 5μl | 1000 | 0.60% | 6 | 0.20% | 2 |
500 | 0.70% | 3.5 | 0.25% | 1.25 | ||
200 | 0.90% | 1.8 | 0.30% | 0.6 | ||
1000-5000μl | 50μl | 5000 | 0.50% | 25 | 0.15% | 7.5 |
2500 | 0.60% | 15 | 0.30% | 7.5 | ||
1000 | 0.70% | 7 | 0.30% | 3 | ||
2-10ml | 0.1ml | 10ml | 0.60% | 60 | 0.20% | 20 |
5ml | 1.20% | 60 | 0.30% | 15 | ||
2ml | 3.00% | 60 | 0.60% | 12 |
8 Chnnel adjustable volumen pipette | ||||||
Volume Range | Incrementum | Test Volume(μl) | accurate error | Subtilitas errorum | ||
% | μl | % | μl | |||
0.5-10μl | 0.1μl | 10 | 1.50% | 0.15 | 1.50% | 0.15 |
5 | 2.50% | 0.125 | 2.50% | 0.125 | ||
1 | 4.00% | 0.04 | 4.00% | 0.04 | ||
5-50μl | 0.5μl | 50 | 1.00% | 0.5 | 0.50% | 0.25 |
25 | 1.50% | 0.375 | 1.00% | 0.25 | ||
5 | 3.00% | 0.15 | 2.00% | 0.1 | ||
50-300μl | 5μl | 300 | 0.70% | 2.1 | 0.25% | 0.75 |
150 | 1.00% | 1.5 | 0.50% | 0.75 | ||
50 | 1.50% | 0.75 | 0.80% | 0.4 | ||
12 Chnnel adjustable volumen pipette | ||||||
Volume Range | Incrementum | Test Volume(μl) | accurate error | Subtilitas errorum | ||
% | μl | % | μl | |||
0.5-10μl | 0.1μl | 10 | 1.50% | 0.15 | 1.50% | 0.15 |
5 | 2.50% | 0.125 | 2.50% | 0.125 | ||
1 | 4.00% | 0.04 | 4.00% | 0.04 | ||
5-50μl | 0.5μl | 50 | 1.00% | 0.5 | 0.50% | 0.25 |
25 | 1.50% | 0.375 | 1.00% | 0.25 | ||
5 | 3.00% | 0.15 | 2.00% | 0.1 | ||
50-300μl | 5μl | 300 | 0.70% | 2.1 | 0.25% | 0.75 |
150 | 1.00% | 1.5 | 0.50% | 0.75 | ||
50 | 1.50% | 0.75 | 0.80% | 0.4 |
Fixum volumen pipettes
Volume Range | Incrementum | Test Volume(μl) | accurate error | Subtilitas errorum | ||
% | μl | % | μl | |||
5μl | - | 5μl | 1.3% | 0.093 | 1.2% | 0.06 |
10μl | - | 10μl | 0.8% | 0.08 | 0.8% | 0.08 |
20μl | - | 20μl | 0.6% | 0.12 | 0.5% | 0.1 |
25μl | - | 25μl | 0.5% | 0.125 | 0.3% | 0.093 |
50μl | - | 50μl | 0.5% | 0.25 | 0.3% | 0.15 |
100μl | - | 100μl | 0.5% | 0.5 | 0.3% | 0.3 |
200μl | - | 200μl | 0.4% | 0.8 | 0.2% | 0.4 |
250μl | -- | 250μl | 0.4% | 1.0 | 0.2% | 0.5 |
500μl | - | 500μl | 0.3% | 1.5 | 0.2% | 1.0 |
1000μl | - | 1000μl | 0.3% | 3.0 | 0.2% | 2.0 |
2000μl | - | 2000μl | 0.3% | 6.0 | 0.15% | 3.0 |
5000μl | - | 5000μl | 0.3% | 15 | 0.15% | 7.5 |